Volume 17, Number 83 (11-2009)                   daneshvarmed 2009, 17(83): 43-48 | Back to browse issues page


XML Persian Abstract Print


Download citation:
BibTeX | RIS | EndNote | Medlars | ProCite | Reference Manager | RefWorks
Send citation to:

mousavi S A J, niakan lahiji M, akhovatian F, moradi moghadam O, valizade hassanlouei M A. An Investigation of endotracheal Tube cuff Pressure. daneshvarmed. 2009; 17 (83) :43-48
URL: http://daneshvarmed.shahed.ac.ir/article-1-39-en.html

Professor
Abstract:   (9112 Views)

  Background and Objective: The objective of this study was to investigate the endotracheal tube cuff (ETT) pressure in the intensive care unit (ICU) patients, who had been intubated.

  

  Meterials and Methods: All patients were placed in ICU during the last three month 2008. The endotracheal tube cuff (ETT) pressure in 308 cases was measured at twice by an expert. A standard manometer was used to measure the cuff pressure and it was calibrated before and after the measurement.

  

  Results: The results show indicated that 50% of cases 1 suffered from brain injuries and 63.6% of them were intubated as a result of brain problem. The value of cuff pressure was not standard in 31 cases at both times. In addition, the cuff pressure in 49% of cases was not standard at first, but after correction it decreased to 18.5%.This emphasizes the effectiveness of having supervision over cuff pressure in acceptable regular intervals by treatment team.

  

  Conclusion: It is suggested that along with recording the vital signs of the patieut, ETT cuff pressure beingrecorded at appropriate intervals as well.

Full-Text [PDF 196 kb]   (1862 Downloads)    
Type of Study: Research |

Add your comments about this article : Your username or email:
Write the security code in the box

Send email to the article author


© 2015 All Rights Reserved | Daneshvar

Designed & Developed by : Yektaweb